RSA-TDP Differential Pressure Transmitter

Applications & Features

  • Apply high accuracy MEMS sensor and digital technologies, can measure positive, negative or differential pressure
  • It can measure system pressure of fan, blower, filter, furnace draft and orifice plate and can apply to various clean room, biological safety cabinet, clean bench, ducts collection, medical or pharmaceutical machine, etc.
  • Multiple ranges, engineering units and outputs
  • Good performance with accuracy of 1.0% and range as low as 25Pa
  • Function keys: zero calibrate, unit select, response time set, etc
  • Field upgradable LCD display module and multiple ranges selection

Specifications

Medium: on-combustible, non-corrosive air, insensitive to moisture, dust, condensation and oil
Working Temp.: -20~70°C
Medium Temp.:0~60°C
Temp. Compensation: 0~50°C
Working Pressure: overload 10xFS, burst 15xFS Accuracy: ±1.0%FS(±2.0%FS@25Pa range)
Long term stability: ±0.5%FS/Year
Thermal effect: <0.05%FS/°C (zero), <0.08%FS/°C (FS) Response Time: 0.5~30s, can be set by keys
Process Connection: 5mm ID tubing
Display: 5 digits LCD, with unit indication, field upgradable Output: 0~10V, 4~20mA (2 wires), RS485 selectable Output Load: ≤500Ω (current), ≥2KΩ (voltage)
Power: Current: 18.5~35VDC (RL=500Ω); 8.5~35VDC (RL=0Ω) Voltage: 16~28VAC/ 16~35VDC
Unit select: by keys

Specifications

Medium: on-combustible, non-corrosive air, insensitive to moisture, dust, condensation and oil
Working Temp.: -20~70°C
Medium Temp.:0~60°C
Temp. Compensation: 0~50°C
Working Pressure: overload 10xFS, burst 15xFS Accuracy: ±1.0%FS(±2.0%FS@25Pa range)
Long term stability: ±0.5%FS/Year
Thermal effect: <0.05%FS/°C (zero), <0.08%FS/°C (FS) Response Time: 0.5~30s, can be set by keys
Process Connection: 5mm ID tubing
Display: 5 digits LCD, with unit indication, field upgradable Output: 0~10V, 4~20mA (2 wires), RS485 selectable Output Load: ≤500Ω (current), ≥2KΩ (voltage)
Power: Current: 18.5~35VDC (RL=500Ω); 8.5~35VDC (RL=0Ω) Voltage: 16~28VAC/ 16~35VDC
Unit select: by keys

Models




Measuring Rangers

  1. Code 1 and 3 have multiple ranges which could be jumper selected.
  2. Set the 5 engineering units by button keys and the related LCD indicator will be on.
  3. For zero center models, add “Z” at the end of the model. For example, RSA-TDP1**Z, means the range is-30-0-30/-62.5-0-62.5pa. Only ranges 1~6 have this selection.